P4.3 Calibration buffer selection sets

Available in pH 310 meter only

The pH 310 meter lets you select between three standard
calibration buffer sets, depending on your requirements.
The available sets are USA, NIST, and DIN standard
calibration buffers.

From measurement mode

1.

Press SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P4.0.

3. Press the ENTER key three times to select

parameter 4.3.

4.

Press the MI/▲ or MR/▼ keys to select the buffer
set you require:

• USA buffers – pH 1.68, 4.01, 7.00, 10.01, and

12.45.

• NIST buffers – pH 1.68, 4.01, 6.86, 9.18, and

12.45.

• DIN buffers – pH 1.09, 3.06, 4.65, 6.79, 9.23

and 12.75.

5. Press the ENTER key to confirm selection and to

return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.

See Figure 31.

Figure 31 - P4.3: Select calibration

buffer sets (pH 310 meter only)