P4.2: Selecting number of pH calibration points

Program P4.2 lets you select the number of calibration
points that appear in pH calibration mode: 2, 3, 4, or 5.
This lets the meter scroll through the calibration points
more quickly if you regularly calibrate at less than 5
points.

From measurement mode.

1.

Press SETUP key to enter Set Up mode.

2. Press the MI/▲ or MR/▼ keys to scroll through

subgroups until you view parameter P4.0.

3. Press the ENTER key twice to select parameter

4.2.

4.

Press the MI/▲ or MR/▼ keys to select 2, 3, 4, or
5 point pH calibration (up to 6 points for DIN
calibration set with pH 310 meter).

5.

Press the ENTER key to confirm selection and to
return to the subgroup menu. Press CAL/MEAS
key to return to measurement mode.

See Figure 30.

Figure 30 - P4.2: Select number of pH

calibration points