beautypg.com

Thermo Fisher Scientific CyberScan pH 300/310 User Manual

Page 45

background image

42

1 1 S P E C I F I CAT I O N S

SPECIFICATIONS DESCRIPTIONS

pH

300

pH

310

pH Range

-2.00 to 16.00 pH

Yes

Yes

Resolution / Accuracy

0.01 pH /

± 0.01 pH

Yes Yes

Slope Range

80 to 110%

Yes

Yes

Temperature Range

-10 to 110

°C

Yes Yes

(14 to 230

°F)

No Yes

Resolution / Accuracy

0.1

°C / ± 0.5 °C (0.1 °F / 0.5 °F)

Yes Yes

Millivolt Range (Extended)

-199.9 to +199.9 mV

Yes

Yes

Resolution / Accuracy

0.1 mV /

± 0.2 mV

Yes Yes

Millivolt Range

-1999 to +1999 mV

Yes

Yes

Resolution / Accuracy

1 mV /

± 2 mV

Yes Yes

Temperature Compensation

Auto / Manual (from 0 to 100

°C)

Yes Yes

Number of calibration points

1 to 5

(up to 6 for pH 310 using DIN)

Yes Yes

Number of calibration buffers

USA: 1.68, 4.01, 7.00, 10.01, & 12.45

Yes

Yes

NIST: 1.68, 4.01, 6.86, 9.18, & 12.45

No Yes

DIN: 1.09, 3.06, 4.65, 6.79, 9.23, &

12.75

No Yes

pH Slope Display

80 to 110%

Yes

Yes

Relative mV & Offset Display

Offset up to +/-150 mV

Yes

Yes

Hold Function

Yes

Yes

Auto-hold

No Yes

Memory

pH or mV plus temperature

16

50

Date and Time Stamping

No Yes

Auto Power off

20 minutes after last key operation

Yes

Yes

Averaging/Stability (READY)

Selectable

Yes

Yes

Inputs

BNC & 6-pin military type

Yes

Yes

Display Custom

Dual

LCD

Yes

Yes

Power Requirements

4 ‘AAA’ Batteries

Yes

Yes

Battery Life

> 200 hours

Yes

Yes

Dimension / Weight

Meter: 19 x 10 x 6 cm; 320 g

Boxed: 24 x 23 x 7 cm; 700 g

Yes Yes