pH 310 meter setup mode

P1.0: Memory clear

P1.0 Clear all stored readings

P2.0: Viewing previous calibration data (USA, NIST

or DIN standards)

P2.1

First calibration point (plus time and date)

P2.2

Second calibration point (plus time and date)

P2.3

Third calibration point (plus time and date)

P2.4

Fourth calibration point (plus time and date)

P2.5

Fifth calibration point (plus time and date)

P2.6

Sixth calibration point (plus time and date)

(only for DIN buffer set)

P3.0: Viewing electrode data

In pH measurement mode:

P3.1

pH electrode offset

P3.2

pH electrode slope

In mV (or relative mV) measurement mode:

P3.1

Relative mV offset

P4.0: Unit configuration

P4.1

READY indicator on or off

P4.2

Number of calibration points: 2, 3, 4 or 5 (or
6 for DIN buffer set only)

P4.3

Calibration buffer selection sets

P4.4

Selecting °C or °F

P5.0: Setting Clock

• Setting

year

• Setting

date

(month/day)

• Setting

time

(hour/minute/second)

P6.0: Reset to factory default

P6.0

Reset to factory default settings