3

Tension Measurement

To measure tensile force, use the hook attachment. The

display will show the force measured and a minus sign (-

) will appear on the left of the display (to the left of the

digits).

NOTE: To display no sign (plus) for tensile force, consult

the “Function Setting” section.

Compression Measurement

To measure compression force, use the flat head

attachment. No sign will appear on the left of the display

during a compression measurement.

NOTE: To display minus sign (-) for compression force,

consult the “Function Setting” section.

Low Battery

When battery charge is low, LO BAT will appear on the

display indicating the batteries need to be charged.

Charging time of fully depleted batteries is approximately

18 hours when the unit is off. The adapter/charger

automatically shuts off when the battery is at full charge

to protect the battery.

Note: The adapter/charger can be used to power the unit

during battery charging, but will lengthen charging time.

Auto Power Off

If the gauge is on and there is no activity for 10 minutes,

the unit automatically powers off to conserve battery

charge. PWR appears above the display digits to notify

that there is 1 minute before power off. If the adapter/

charger is powering the gauge, auto power off function

becomes inactive.

NOTE: To modify the auto power off function, consult

the “Function Setting” section.

Tracking Function

A tracking function has been introduced to check and

compensate for temperature drift. When measuring very

minute forces (a few ounces or grams) and at a very

slow rate, you may want to turn off the tracking function.

To turn the tracking function off, follow the steps below:
1. Turn POWER off.
2. Press PEAK and UNIT switches simultaneously and

hold, then press and release POWER (If tracking

function was on, the display will show TRK OFF

momentarily)

3. Release PEAK and UNIT.
The tracking function is now off. Repeat above steps to

turn it on. The gauge will display TRK SET when turned

on. It is a good idea to have this function on at all the

times unless it is absolutely necessary to cancel it.

External Device Communications

When operated with the appropriate output cable, the

FGV-X offers output capabilities.

Overload

Even though each model is able to withstand an overload

of 200% of its rated capacity, caution should be exercised

that this does not happen, otherwise the sensor will be

damaged. To protect the gauge and/or the sample under

test when a motorized stand or some other device is

used in conjunction with the gauge, two overload output

OC NPN transistors are available to be used to disconnect

power when the overload condition reaches 120% of

the gauge’s rated capacity. One transistor is for tension

and the other for compression. See figure 2.

Analog Output

An analog output signal is available for recording

purposes. The amplitude of this signal is ± 1 VDC. The

voltage is positive when compression testing is performed

and negative for tension. The signal characteristics are

as follows:

Amplitude

Amplitude

Amplitude

Amplitude

Amplitude

± 1 VDC

Generated by

Generated by

Generated by

Generated by

Generated by

12-bit D/A converter

Signal update

Signal update

Signal update

Signal update

Signal update

100 times/sec (update every 10 msec)

Load impedance

Load impedance

Load impedance

Load impedance

Load impedance

10 KW minimum

Connector pins

Connector pins

Connector pins

Connector pins

Connector pins

Pin #1

signal output (analog)

Pin #2

GND (analog)

(see Fig.3 & Table 1)

NOTE: When the zero switch is pressed to tare the gauge

the analog output goes to 0V automatically.

Fig. 2