Page 8: Error codes
Mode 5: Self T
Mode 5: Self T
Mode 5: Self T
Mode 5: Self T
Mode 5: Self Test
est
est
est
est
This mode lets you check the LED display and input
circuitry.
1. After disconnecting sensor, press
mode and data set keys for five
seconds.
2. Press increment key until
display
Specifications
Specifications
Specifications
Specifications
Specifications
Model
DT-5TP (VDC)
Mode
Rate Measurement
Display Range
0.0000–9.9999
0.000–99.999
0.00–999.99
0.0–9999.9
0–99999
Measuring range
10–99999 rpm (at 1p/r), 0.2–30000 rpm
(at 60p/r)
Display period
1 sec. (1 p/r input 60 rpm or over)
1–6 sec. (1 p/r input 10–60 rpm)
Display
Red LED (0.56" or 14.2 mm high)
Time base
Controlled by a 4.194304 MHz crystal
Accuracy
±0.008% ± 1 digit
Measuring system
CPU controlled
Input no. of p/r
1–9999 (programmable)
Input signal
characteristics
Sine wave–max frequency 10 KHz
Square wave–max frequency 30 KHz
Open collector–max frequency 30 KHz
Contact closure–max frequency 20 Hz
Input signal amplitude
Sine wave (0.3–30 VP–P)
Square wave LO: 0–1.5 V, HI 4–30 V
Input impedance
Approx. 10 k ohms
Voltage output
12 VDC ±5% (50 mA max) to power sensors
Ambient temperature
32°–113° F (0°–45° C)
Power consumption
1W
Voltage requirements
9–35 VDC
Dimensions
3.46"L x 1.88"H x 3.78"W (88L x 48H x 96W
mm), includes bezel, fit 1/8 DIN cutout
Weight
0.55 lbs (250g)
Error Codes
Error Codes
Error Codes
Error Codes
Error Codes
Display
Type of Error
What to Do
EE-02
Internal memory
Press data set key. Interrupt power at terminals 1 and 2.
EE-03
Memory recall
1. Press data set key.
2a. Press increment key. Display should show
2b. Press data set key. Display should show
2c. Press shift key. Display should show
2d. Press mode key. Display should show
and reset to
. If display goes to
or
and
locks up, turn power off and on again and repeat above steps.
SHIMPO INSTRUMENTS
1701 Glenlake Avenue Itasca, IL 60143 USA (630) 924-7138 Fax (630) 924-0342
shows mode 5.
3. Press mode key to test display
segments and decimal points.
4. Press mode key. The display will
show
and the signal
will flash.
5. Press the data set key. The unit
is
now ready for counting.