Mode 5: Self T

Mode 5: Self T

Mode 5: Self T

Mode 5: Self T

Mode 5: Self Test

est

est

est

est

This mode lets you check the LED display and input
circuitry.

1. After disconnecting sensor, press

mode and data set keys for five
seconds.

2. Press increment key until

display

Specifications

Specifications

Specifications

Specifications

Specifications

Model

DT-5TP (VDC)

Mode

Rate Measurement

Display Range

0.0000–9.9999
0.000–99.999
0.00–999.99
0.0–9999.9

0–99999

Measuring range

10–99999 rpm (at 1p/r), 0.2–30000 rpm

(at 60p/r)

Display period

1 sec. (1 p/r input 60 rpm or over)

1–6 sec. (1 p/r input 10–60 rpm)

Display

Red LED (0.56" or 14.2 mm high)

Time base

Controlled by a 4.194304 MHz crystal

Accuracy

±0.008% ± 1 digit

Measuring system

CPU controlled

Input no. of p/r

1–9999 (programmable)

Input signal
characteristics

Sine wave–max frequency 10 KHz
Square wave–max frequency 30 KHz

Open collector–max frequency 30 KHz
Contact closure–max frequency 20 Hz

Input signal amplitude

Sine wave (0.3–30 VP–P)
Square wave LO: 0–1.5 V, HI 4–30 V

Input impedance

Approx. 10 k ohms

Voltage output

12 VDC ±5% (50 mA max) to power sensors

Ambient temperature

32°–113° F (0°–45° C)

Power consumption

1W

Voltage requirements

9–35 VDC

Dimensions

3.46"L x 1.88"H x 3.78"W (88L x 48H x 96W
mm), includes bezel, fit 1/8 DIN cutout

Weight

0.55 lbs (250g)

Error Codes

Error Codes

Error Codes

Error Codes

Error Codes

Display

Type of Error

What to Do

EE-02

Internal memory

Press data set key. Interrupt power at terminals 1 and 2.

EE-03

Memory recall

1. Press data set key.
2a. Press increment key. Display should show

2b. Press data set key. Display should show

2c. Press shift key. Display should show

2d. Press mode key. Display should show

and reset to

. If display goes to

or

and

locks up, turn power off and on again and repeat above steps.

SHIMPO INSTRUMENTS

1701 Glenlake Avenue Itasca, IL 60143 USA (630) 924-7138 Fax (630) 924-0342

shows mode 5.

3. Press mode key to test display

segments and decimal points.

4. Press mode key. The display will

show

and the signal

will flash.

5. Press the data set key. The unit

is

now ready for counting.