8.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Faraday's law
Application range
Electrically conductive fluids
Measured value
Measured value
Measured value
Measured value
Primary measured value
Flow velocity
Secondary measured value
Volume flow
Design
Modular construction
The measurement system consists of a flow sensor and a signal
converter. It is available as compact and as remote version. More
information about the sensor can be found in the relevant
documentation.
Compact version
With WATERFLUX 3000 sensor: WATERFLUX 3070 C
With OPTIFLUX 2000 sensor: OPTIFLUX 2070 C
Remote version
With WATERFLUX 3000 sensor: WATERFLUX 3070 F
With OPTIFLUX 2000 sensor: OPTIFLUX 2070 F
Maximum cable length 25 m / 75 ft
User interface
User interface
User interface
User interface
Display
8 digits LCD
Display of positive and negative counter, sum counter, flow rate
Status indication for battery, flow / counter direction, empty pipe
Units
Volume in m
3
, US Gallons
Flow rate in m
3
/h, USGPM, l/s
Cable connections
Standard: 2x M20x1.5
Optional: ½" NPT, PF½
Measuring accuracy
Reference conditions
Medium: water
Temperature: 20°C / 68°F
Inlet section: 5 DN
Operating pressure: 1 bar / 14.5 psig
Maximum measuring error
±
0.2% of measured value ±0.5 mm/s / 0.02 inch/s
For detailed information on the measuring accuracy, see chapter
"Measuring accuracy".
Repeatability
±
0.1% (v > 0.5 m/s / 1.5 ft/s)