beautypg.com

Operation, 4 scaling – KROHNE OPTISENS OAS 2000 EN User Manual

Page 36

background image

5

OPERATION

36

OPTISENS OAS 2000

www.krohne.com

07/2010 - MA OAS 2000 R02 en

Calibration of submersible sensor in a basin or channel

• Calibration can be done without the use of a bucket. Make sure that the sensor is at least

30 cm / 11.8" below the lowest liquid level.

• Follow the steps 3 to 5 of the above procedure "Calibration with a sample bucket".
• While the OAS 2000 is calibrating, grab a sample of the liquid with a dip bucket. Make sure to

grab a sufficiently large sample volume for low solids applications.

• Take the sample to the lab for analysis. Note the concentration of the sample determined at

the lab.

• Follow the steps 8 to 10 of the above procedure "Calibration with a sample bucket".

Calibration of inline sensor in a pipe

• Inline sensors are easiest calibrated if the pipe has a sample outlet.
• Follow the steps 3 to 5 of the above procedure "Calibration with a sample bucket".
• While the OAS 2000 is calibrating, open the sample valve and fill a bucket with process liquid.
• Take the sample to the lab for analysis. Note the concentration of the sample determined at

the lab.

• Follow the steps 8 to 10 of the above procedure "Calibration with a sample bucket".

5.4 Scaling

The menu Scale / Alarm

Scale / Alarm

Scale / Alarm

Scale / Alarm (see converter manual) allows the user to set the high and low

boundaries for a 4...20 mA output signal. In addition, this menu allows the user to set high and
low alarm values to switch a relay when solids have reached critical points.

Max

Max

Max

Max

Sets the 20 mA point output.

Min

Min

Min

Min

Sets the 4 mA point output (may be negative for special applications).

Hi-Alarm

Hi-Alarm

Hi-Alarm

Hi-Alarm

Sets the high alarm set point (the value zero inactivates the alarm).

Low-Alarm

Low-Alarm

Low-Alarm

Low-Alarm

Sets the low alarm set point (the value zero inactivates the alarm).

.book Page 36 Friday, July 2, 2010 1:49 PM