Operation

5.1 Sensor display

By simultaneously pressing ↓ and ^ you alter between the converter main menu and the sensor
information display for the selected sensor. The sensor information display shows the sensor's
calibration curve.

5.2 Menu for OAS 2000 sensor

Use ↑ or ↓ to select the sensor in the main display. The menu for the selected sensor is accessed
by pressing ^ for five seconds. If the selected sensor is not active (the text No transmitter

No transmitter

No transmitter

No transmitter is

shown) a warning is displayed that asks you to make another choice in order to show the sensor
menu.

Menu "Settings"

Menu "Calibrate"

Submenu

Description

Tag

Tag

Tag

Tag

Name of the sensor shown in the main display (10 characters).

I-Time

I-Time

I-Time

I-Time

Integration time or dampening (can be set up to 999 seconds).

Unit

Unit

Unit

Unit

"%", "ppm", "g/l", "mg/", NTU, FNU

Decimals

Decimals

Decimals

Decimals

"Std" or "Extra". Number of decimals for the reading.

Analog

Analog

Analog

Analog

"None" , "Out1", "Out2", "Out3", "Out4", "Out1+2" or "Out3+4". Pick

which analog output(s) should be used with sensor.

Second

Second

Second

Second

"Temp" or "=Prim". If two analog outputs are chosen above, the first

will always give the primary value according to the sensors selected

scale. The second will either give the temperature scaled 0...100°C /

32...212°F, or the same signal as the first channel. The temperature

is additional information, not a precision measurement.

Submenu

Description

Adjust

Adjust

Adjust

Adjust

"No", "Store" or "Lab". Stores the reading of the meter when a

sample is taken and can then automatically adjust the sample value

when the sample analysed by lab is different to the reading.

Take sample

Take sample

Take sample

Take sample

"No", "Zero", "1", "2", "3", "4", "5". Sensor stores current MS (light)

value in memory and you enter a lab solids value below to complete

calibration.

Con

Con

Con

Con

Current concentration (same as shown in the main menu).

Sample 1

Sample 1

Sample 1

Sample 1

Lab test (suspended solids value for Sample 1)

Sample 2

Sample 2

Sample 2

Sample 2

Lab test (suspended solids value for Sample 2)

Sample 3

Sample 3

Sample 3

Sample 3

Lab test (suspended solids value for Sample 3)

Sample 4

Sample 4

Sample 4

Sample 4

Lab test (suspended solids value for Sample 4)

Sample 5

Sample 5

Sample 5

Sample 5

Lab test (suspended solids value for Sample 5)