Operation – KROHNE OPTISENS OAS 2000 EN User Manual
Page 35
OPERATION
5
35
OPTISENS OAS 2000
www.krohne.com
07/2010 - MA OAS 2000 R02 en
Running a zero calibration
• Remove the sensor from the process and clean the sensor head.
• Dip the sensor in a bucket of clean water.
• Select the sensor to be calibrated in the menu by using ↑ or ↓.
• Press ^ for approximately 5 seconds to enter the sensor menu.
• Select Calibrate > Take sample
Calibrate > Take sample
Calibrate > Take sample
Calibrate > Take sample and then press ^.
• Select Zero
Zero
Zero
Zero by using ↑ or ↓ and then press ^.
• To acknowledge that you really want to change the zero calibration, select Yes
Yes
Yes
Yes and then press
^.
i
The converter will ask you to put the sensor in clean water.
• Submerge the sensor head into the clean water and cover it from direct sunlight.
• Press ^.
• Wait for the zero calibration to finish. It will take approximately 30 seconds before the unit
returns to the menu.
5.3.7 Calibration with samples
Calibration with sample in a bucket
• Fill a bucket with a sample of the liquid you intend to measure.
• Submerge the sensor into the liquid.
• Select the sensor to be calibrated in the menu by using ↑ or ↓.
• Press ^ for approximately five seconds to enter the sensor menu.
• Select Calibrate > Take sample
Calibrate > Take sample
Calibrate > Take sample
Calibrate > Take sample and then #1
#1
#1
#1 using ↑ and ↓.
• Press ^ and stir the sample in the bucket until the calibration is finished. It will take
approximately 30 seconds.
• Take the bucket to the lab for analysis. Note the concentration of the sample determined at
the lab.
• Enter sample #1 concentration by selecting Calibrate > Sample #1
Calibrate > Sample #1
Calibrate > Sample #1
Calibrate > Sample #1 in the calibration menu.
• Press ^.
• Use ↑ and ↓ to change the values and ^ to go to the next digit. Some special applications may
need additional sample points. Do not enter samples that are identical in concentration or
less than 10% from initial values.
INFORMATION!
Detailed procedures for navigating the converter software can be found in the converter manual.
CAUTION!
Even though the sensors have daylight-filters, they are sensitive to the infrared parts of the
sunlight. Always cover the sensor and the bucket before calibration.
.book Page 35 Friday, July 2, 2010 1:49 PM