beautypg.com

KROHNE SU 600 EN User Manual

Page 31

background image

Relay outputs
Quantity

2x operating relay, 1x fail safe relay

Contact

floating spdt

Contact material

AG NI 0.15 hard gold-plated

Turn-on voltage

min. 10 mV DC, max. 250 V AC/DC

Switching current

min. 10 µA DC, max. 3 A AC, 1 A DC

Breaking capacitance

min 50 mW, max. 500 VA, max. 54 W DC

Min. switching hysteresis

0.5 %

Current outputs
Quantity

1x output

Range

0/4 … 20 mA

Resolution

0.1 % or 20 µA

Max. load

500 Ohm

Fault signal

22 mA

Linearity error

0.1 % (relating to 20 mA)

Temperature error

0.01 %/K (relating to 20 mA)

Displays
Measured value indication
- LC display (45x45 mm)

digital and quasianalogue indication

- max. indicating range

-9999 ... 9999

LED displays
- status indication operating vol-

tage

1x LED green

- status indication fault signal

1x LED red

- status indication operating re-

lay 1/2

2x LED yellow

Adjustment
Adjustment elements

2x keys in the front, 1x function switch

Ambient conditions
Ambient temperature

-20 … +60°C (-4 … +140°F)

Storage and transport temperature

-40 … +80°C (-40 … +176°F)

SU 600 - 4 … 20 mA signal conditioning

31

Supplement

30652

-EN

-050525