KROHNE SU 600 EN User Manual
Page 31
Relay outputs
Quantity
2x operating relay, 1x fail safe relay
Contact
floating spdt
Contact material
AG NI 0.15 hard gold-plated
Turn-on voltage
min. 10 mV DC, max. 250 V AC/DC
Switching current
min. 10 µA DC, max. 3 A AC, 1 A DC
Breaking capacitance
min 50 mW, max. 500 VA, max. 54 W DC
Min. switching hysteresis
0.5 %
Current outputs
Quantity
1x output
Range
0/4 … 20 mA
Resolution
0.1 % or 20 µA
Max. load
500 Ohm
Fault signal
22 mA
Linearity error
0.1 % (relating to 20 mA)
Temperature error
0.01 %/K (relating to 20 mA)
Displays
Measured value indication
- LC display (45x45 mm)
digital and quasianalogue indication
- max. indicating range
-9999 ... 9999
LED displays
- status indication operating vol-
tage
1x LED green
- status indication fault signal
1x LED red
- status indication operating re-
lay 1/2
2x LED yellow
Adjustment
Adjustment elements
2x keys in the front, 1x function switch
Ambient conditions
Ambient temperature
-20 … +60°C (-4 … +140°F)
Storage and transport temperature
-40 … +80°C (-40 … +176°F)
SU 600 - 4 … 20 mA signal conditioning
31
Supplement
30652
-EN
-050525
- BATCHFLUX 5500 C Quickstart EN (20 pages)
- IFC 050 Converter Quickstart EN (28 pages)
- IFC 100 Converter Quickstart EN (32 pages)
- IFC 300 Converter Quickstart EN (68 pages)
- OPTIFLUX 1000 Quickstart EN (20 pages)
- OPTIFLUX 2000 Quickstart EN (24 pages)
- OPTIFLUX 4000 Quickstart EN (24 pages)
- OPTIFLUX 4040C Quickstart EN (16 pages)
- OPTIFLUX 5000 Flange Quickstart EN (20 pages)
- OPTIFLUX 5000 Sandwich Quickstart EN (20 pages)
- OPTIFLUX 6000 Quickstart EN (28 pages)
- OPTIFLUX 7300 Quickstart EN (24 pages)
- OPTIPROBE Quickstart EN (16 pages)
- TIDALFLUX 2300 F EN (44 pages)
- TIDALFLUX 2300 F Quickstart EN (24 pages)
- WATERFLUX 3000 EN (40 pages)
- WATERFLUX 3000 Quickstart EN (24 pages)
- WATERFLUX 3070 EN (80 pages)
- WATERFLUX 3070 Quickstart EN (32 pages)
- USB ADAPTER PLUS EMF EN (16 pages)
- IFC 050 Converter Modbus EN (20 pages)
- IFC 100 Converter FOUNDATION FIELDBUS EN (64 pages)
- IFC 100 Converter Modbus EN (20 pages)
- IFC 300 Converter FOUNDATION FIELDBUS EN (60 pages)
- IFC 300 Converter HART 0102 EN (20 pages)
- IFC 300 Converter HART 0201 EN (23 pages)
- IFC 300 Converter Modbus EN (24 pages)
- IFC 300 Converter PROFIBUS PA DP EN (40 pages)
- OPTIFLUX 2000-4000 IECEx EN (16 pages)
- OPTIFLUX 2000-4000-5000-6000-7300-IFC 300 Ex EN (40 pages)
- OPTIFLUX 2000-4000-5000-6000 -IFC 100 Ex EN (24 pages)
- OPTIFLUX 4040 C Ex EN (20 pages)
- OPTIFLUX x300 Ex Zone2 EN (1 page)
- H250 M9 ES EN (36 pages)
- VA 40-VA 45 EN (36 pages)
- H250 M10 ATEX II2G Ex d EN (16 pages)
- H250 M10 ATEX II3D Ex t EN (16 pages)
- H250 M40 ATEX II2D Ex t-II2G Ex d EN (20 pages)
- H250 M40 ATEX II2G Ex i EN (20 pages)
- H250 M40 ATEX II3G Ex nA EN (20 pages)
- H250 M40 Ex II2G Reed EN (2 pages)
- H250 M9 ATEX II2G Ex i EN (16 pages)
- H250 M9S ATEX II3D Ex t-II3G Ex nA EN (20 pages)
- M8E Converter HART 0101 EN (13 pages)
- DK 32-DK 34 ATEX II2G Ex i EN (16 pages)