E016

l

Empty/full adjustment reversed

à

Carry out a fresh adjustment

?

E017

l

Adjustment span too small

à

Carry out a fresh adjustment and increase the

distance between min. and max. adjustment

?

E021

l

Scaling span too small

à

Carry out a fresh scaling, increase the distance

between min. and max. scaling.

?

E110

l

Relay switching points too close together

à

Increase the difference between the two relay

switching points

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form from our Internet

homepage http://www.krohne-mar.com/fileadmin/

media-lounge/PDF-Download/Specimen_e.pdf.
By doing this you help us carry out the repair quickly and

without having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and possibly also a safety

data sheet to the instrument.

28

SU 600 - 4 … 20 mA signal conditioning

Maintenance and fault rectification