beautypg.com

KROHNE SU 600 EN User Manual

Page 26

background image

Because the range of the example vessel is between 50 l

to 2650 l, you will need all four positions of the digital

indication.
1 Set the function switch to position [5] (Decimal Point).
2 By pushing the [+/-] keys, you move the decimal

point.

3 Save the value by pushing both keys simultaneously.

l

Set the function switch to position [ 1 ] (relay 1 on).

For the given example, select the value 90.0 (90.0 %)

with the [ +/- ] keys. The internal relay 1 then switches

on when this value is reached. Save the value by

pushing the two keys simultaneously.

l

Set the function switch to position [ 2] (relay 1 off).

Select the value 10.0 (10.0 %) with the [ +/-] keys.

The internal relay 1 then switches off when the value

falls below this value. The respective relay control

lamp lights when the relay is energized. If the

switching points are too close together (<0.1 %), the

appropriate LED flashes. The relay takes on the safe

condition. In mode "OPERATE " an error message is

outputted on the display.

l

If you want to control additional relays, proceed in the

same way, as with relay 1. The switch positions for

setting the respective relays are stated in the list of

the function switch under "Setup procedure".

Note:

If you want to change the mode (i.e. the switching

function of the relays), you have to exchange the On and

Off values.

Decimal point

Relay

26

SU 600 - 4 … 20 mA signal conditioning

Setup with the integrated indicating and adjustment unit

30652

-EN
-050525