7 Maintenance and fault rectification

7.1 Maintenance

When used as directed in normal operation, SU 600 is

completely maintenance-free.

7.2 Fault rectification

SU 600 offers maximum reliability. Nevertheless faults

can occur during operation. These may be caused by

the following, e.g.:

l

Measured value of the sensor not correct

l

Power supply

l

Interference on the cables

The first measures to be taken are to check the input/

output signals as well as to evaluate the error messages

via the display. The procedure is described below. In

many cases, the causes can be determined in this way

and faults can be rectified.

?

E003

l

CRC-error

à

Carry out a reset

à

Return instrument for repair

?

E014

l

Short-circuit

à

Eliminate short-circuit

à

If necessary, exchange defective sensor

?

E015

l

Line break

à

Repair sensor cable or replace it

à

Check sensor connection

Causes of malfunction

Fault rectification