5.3 Wiring plan

+ -

1 2

+ -

3 4

18 17 16 15 14 13 12 11 10 9 8 7 6 5

N- L1+

on

1

2

1

2

%

1

2

3

4

5

6

Fig. 5: Terminal assignment SU 600
1 Measurement data input, optionally available with sensor power

supply

2 Current output

3 Fail safe relay

4 Relay 2

5 Relay 1

6 Power supply

Information:

In the menu, it is possible to switch over between active

and passive operation of the measurement data input:

l

In active mode,SU 600 provides the power for the

connected sensors. Power supply and measurement

data are transmitted over the same two-wire cable.

This mode is provided for connection of transmitters

without separate power supply (sensors in two-wire

version).

Overview

16

SU 600 - 4 … 20 mA signal conditioning

Connecting to power supply

30652

-EN
-050525