beautypg.com

Technical data, 2 technical data – KROHNE OPTISENS ORP 8590 EN User Manual

Page 27

background image

TECHNICAL DATA

6

27

OPTISENS ORP 8590

www.krohne.com

01/2014 - 4001933801 MA OPTISENS ORP 8590 R01 en

6.2 Technical data

ORP sensor

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

OPTISENS

ORP 8590

Measuring system

Measuring principle

Potentiometric

Measuring range

-2000…+2000 mV

Design

Construction

CPVC Body mit Tuff Tip®

Shaft diameter

20 mm / 0.79"

Length

23 mm / 0.91" or 44 mm / 1.74"

Process connection

3/4" MNPT

Temperature sensor

Pt100

Sensor cap

4-pin connector or attached cable

Type of diaphragm

Ceramic

Measuring accuracy

Reference conditions

Medium: water

Temperature: 20°C / 68°F

Pressure: max. 1 bar / 14.5 psi (absolute)

Maximum measuring

error

pH: 0.2% full scale

Temperature: 1.0% full scale

Repeatability

0.2% full scale

Resolution

0.1 (or 0.01 in extended mode)

Long-term stability

24 hours: tested within accuracy definition

Temperature drift

Tested within accuracy definition

Cable length variation

Tested within accuracy definition

.book Page 27 Friday, January 24, 2014 11:29 AM