OPERATION

4

21

OPTISENS ORP 8590

www.krohne.com

01/2014 - 4001933801 MA OPTISENS ORP 8590 R01 en

• If you want to return to the measuring mode, press ^

^

^

^ several times until you reach this mode.

Prior returning to the measuring display, you are asked if the configuration should be stored.
Choose yes

yes

yes

yes using

or

to store the new calibration values. Press ^

^

^

^ to confirm.

Step 4: re-installing the sensor

• After the calibration procedure, rinse the sensor with tap water.
• Reinstall the sensor into its assembly, refer to

Installing the sensor on page 15.

Step 5: switching back to measurement

• Deactivate the function hold function

hold function

hold function

hold function again.

4.2.2 Calibration log

Accessing the calibtration log

• You are on the data level and you see the calibration history. With the help of

or

you can

scroll through the different entries.

• If you want to return to the measuring mode press ^

^

^

^ several times until you reach this mode.

INFORMATION!
If an error occurs during the calibration procedure, the display shows an error message.
Possible causes for an error are:

INFORMATION!
In order to show the history of the calibrations, the signal converter has a calibration logbook
function. Up to 64 entries of the calibration history are stored including date and time.

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

Press

or

until the main menu test

test

test

test is highlighted.

MAIN MENU

A quick setup

> B test

> B test

> B test

> B test

C setup

D service

Press >>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "test" and "B1" appears,

beneath the submenu sim.process input A

sim.process input A

sim.process input A

sim.process input A is highlighted.

Press or until the submenu logbooks

logbooks

logbooks

logbooks is highlighted.

Press >>>> to enter the chosen menu.

You are on the second submenu level. In the upper line of the display "logbooks" and

"B1" appears, beneath the submenu status log

status log

status log

status log is highlighted.

Press

or

until the submenu calibration log

calibration log

calibration log

calibration log is highlighted.

Press >

>>> to enter the chosen menu.