Operation – KROHNE OPTISENS ORP 8590 EN User Manual
Page 21
OPERATION
4
21
OPTISENS ORP 8590
www.krohne.com
01/2014 - 4001933801 MA OPTISENS ORP 8590 R01 en
• If you want to return to the measuring mode, press ^
^
^
^ several times until you reach this mode.
Prior returning to the measuring display, you are asked if the configuration should be stored.
Choose yes
yes
yes
yes using
or
to store the new calibration values. Press ^
^
^
^ to confirm.
Step 4: re-installing the sensor
• After the calibration procedure, rinse the sensor with tap water.
• Reinstall the sensor into its assembly, refer to
Installing the sensor on page 15.
Step 5: switching back to measurement
• Deactivate the function hold function
hold function
hold function
hold function again.
4.2.2 Calibration log
Accessing the calibtration log
• You are on the data level and you see the calibration history. With the help of
or
you can
scroll through the different entries.
• If you want to return to the measuring mode press ^
^
^
^ several times until you reach this mode.
INFORMATION!
If an error occurs during the calibration procedure, the display shows an error message.
Possible causes for an error are:
INFORMATION!
In order to show the history of the calibrations, the signal converter has a calibration logbook
function. Up to 64 entries of the calibration history are stored including date and time.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
•
Press
or
until the main menu test
test
test
test is highlighted.
MAIN MENU
A quick setup
> B test
> B test
> B test
> B test
C setup
D service
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "test" and "B1" appears,
beneath the submenu sim.process input A
sim.process input A
sim.process input A
sim.process input A is highlighted.
Press or until the submenu logbooks
logbooks
logbooks
logbooks is highlighted.
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the display "logbooks" and
"B1" appears, beneath the submenu status log
status log
status log
status log is highlighted.
Press
or
until the submenu calibration log
calibration log
calibration log
calibration log is highlighted.
Press >
>>> to enter the chosen menu.
.book Page 21 Friday, January 24, 2014 11:29 AM