4.3 Troubleshooting
Problem
Possible cause
Remedy
The ORP sensor
does not deliver a
signal.
Mechanical damage of the
platinum electrode, e.g.
scratches.
Exchange sensor.
Moisture inside the
sensor connector.
Clean the connector (sensor/cable) with pure water
and dry with air (e.g. hair blower).
The ORP sensor
delivers a bad
signal.
Coated/abraded platinum
electrode.
Clean the platinum electrode by using aqua regia (this
procedure uses hazardous chemicals and should only
be performed by a qualified person), HCl
concentrated/HNO3 concentrated (1/1).
The ORP electrode
delivers an
unstable signal.
The diaphragm in the
reference half-cell does
not provide good contact
to the process medium
due to drying up or
coatings.
•
Clean the diaphragm with hot soap or acid using a
soft tissue (details on page 23).
•
Submerge sensor in water and increase the
temperature to 50...60°C / 122...140°F.
•
Submerge sensor in 3 molar KCl solution at ambient
temperature. The decrease in temperature will cause
the reference half cell to suck in KCl solution through
the diaphragm and regenerate the diaphragms
functionality.