4.3 Troubleshooting

Problem

Possible cause

Remedy

The ORP sensor

does not deliver a

signal.

Mechanical damage of the

platinum electrode, e.g.

scratches.

Exchange sensor.

Moisture inside the

sensor connector.

Clean the connector (sensor/cable) with pure water

and dry with air (e.g. hair blower).

The ORP sensor

delivers a bad

signal.

Coated/abraded platinum

electrode.

Clean the platinum electrode by using aqua regia (this

procedure uses hazardous chemicals and should only

be performed by a qualified person), HCl

concentrated/HNO3 concentrated (1/1).

The ORP electrode

delivers an

unstable signal.

The diaphragm in the

reference half-cell does

not provide good contact

to the process medium

due to drying up or

coatings.

Clean the diaphragm with hot soap or acid using a
soft tissue (details on page 23).

Submerge sensor in water and increase the
temperature to 50...60°C / 122...140°F.

Submerge sensor in 3 molar KCl solution at ambient
temperature. The decrease in temperature will cause
the reference half cell to suck in KCl solution through
the diaphragm and regenerate the diaphragms
functionality.