Installation, 6 calibrating the sensor – KROHNE OPTISENS ORP 8590 EN User Manual

Page 14

background image

3

INSTALLATION

14

OPTISENS ORP 8590

www.krohne.com

01/2014 - 4001933801 MA OPTISENS ORP 8590 R01 en

The following instructions describe the connection of the different sensor cables.

Connecting the sensor cable to the signal converter

• Thread the sensor cable through the outer right cable gland 1.
• Push the coax shield 4 into terminal N 2 and the coax core into terminal O 3.
• To remove a cable, press down the white clip 5 on the corresponding terminal and pull the

cable out 6.

3.5.2 Connecting the cable to the sensor

Connecting the cable to the sensor

• Unscrew the protective cap from the sensor connector and keep it for future use 1.
• Ensure that both cable and sensor connector are absolutely dry 2.
• Make sure that the O-ring is positioned on the sensor connector 3.
• Screw the cable connector 4 on to the sensor and tighten it by hand.

3.6 Calibrating the sensor

Before the sensor is installed, it has to be calibrated. Proceed as described on page 18. Then
continue with the installation procedure.

Figure 3-4: Connecting the sensor cable

CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector, dry it with air (e.g. hair blower).

Figure 3-5: Connecting the cable to the sensor