Operation – KROHNE OPTISENS ORP 8590 EN User Manual

Page 20

background image

4

OPERATION

20

OPTISENS ORP 8590

www.krohne.com

01/2014 - 4001933801 MA OPTISENS ORP 8590 R01 en

Step 1b: accessing the calibration menu via the main menu quick setup

quick setup

quick setup

quick setup

• You can start the calibration procedure now as described in "Step 2".

Step 2: preparing the calibration procedure

• If you re-calibrate an existing sensor, remove the sensor from its respective assembly and.
• If you calibrate a new sensor, make sure that the sensor is correctly connected to the signal

converter.

• Check the sensor for damages, check the diaphragm for coating and rinse the sensor tip with

tap water and gently swipe it with a soft tissue.

• Provide a suitable calibration solution.

After activating the hold function and the preparative measures, you can get access to the
calibration procedure from the measuring mode in two different ways. Either you go via the main
menu setup

setup

setup

setup (step 3a) or via the main menu quick setup

quick setup

quick setup

quick setup (step 3b).

Step 3: calibration procedure

• After choosing the submenu ORP

ORP

ORP

ORP (step 1a) or ORP calibration A or B

ORP calibration A or B

ORP calibration A or B

ORP calibration A or B (step 1b) and step 2 in the

previous steps, continue by pressing >>>>.

i

The currently measured value is shown on the display.

• Press ^

^

^

^ to finally start the calibration procedure.

• Submerge the sensor tip into the ORP buffer solution.
• Wait until a steady value is displayed.
• Press ^

^

^

^ to continue the calibration procedure.

i

The message ref. value ORP

ref. value ORP

ref. value ORP

ref. value ORP and the value of the buffer solution are displayed on the

screen.

• Enter the value of the buffer solution by pressing

or

and >>>>. Press ^

^

^

^ to confirm the

setting.

• After 25 seconds the calibration step is completed.

i

The message zero point is displayed on the screen.

• Press ^

^

^

^ to confirm the setting.

i

The message store cal. value is displayed on the screen. The signal converter asks if the
new calibration values should be stored.

• Choose yes

yes

yes

yes using

or

to store the calibration values. Choose no

no

no

no to discard the results.

• Press ^

^

^

^ to confirm.

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

MAIN MENU

> A quick setup

> A quick setup

> A quick setup

> A quick setup

B test

C setup

D service

Press >>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "quick

setup" and "a" appears, beneath the submenu language

language

language

language is highlighted.

Press

or

until the submenu ORP calibration A

ORP calibration A

ORP calibration A

ORP calibration A or B

B

B

B is highlighted.

Press >>>> to enter the chosen menu.

You can start the calibration procedure now as described in "Step 2".