4.2 Calibration
4.2.1 Calibrating ORP measurement
The potential of an ORP electrode is calibrated using a ORP buffer solution. In the course of that,
the difference between the measured potential and the potential of the calibration solution is
determined. This potential difference is printed on the buffer solution bottle and is defined as the
voltage across the ORP electrode and a reference electrode. ORP sensors require no calibration,
but you can perform a one-point calibration with an ORP buffer solution.
To avoid alarms on the distributed control system (DCS) when temporarily removing the sensor
(i.e. for maintenance), the signal converter has a hold function. This function "freezes" all
outputs (i.e. the display and the current outputs) of the last measured value.
Measuring
mode
Main menu
Submenu
Parameter
3 or 4
pages,
scrolling
with ↓ or ↑
> 2.5 s
^
C setup
C setup
C setup
C setup
>
^
C3 I/O
>
^
C3.1 hardware
>
^
For further
information
see function
tables.
C3.2 current output A
C3.3 current output B
C3.4 current output C
C5 device
>
^
C5.1 device info
C5.2 display
C5.3 1.meas.page
C5.4 2.meas.page
C5.5 graphic page
C5.6 special functions
C5.7 units
↓↑
↓↑
↓↑
↓↑
D service:
D service:
D service:
D service: This menu is password protected and contains functions to be used by service personnel only.
INFORMATION!
Remove the watering cap before starting the calibration.
CAUTION!
•
Never touch or scratch the platinum measuring tip of the ORP sensor.
•
Make sure that the glass tip is clean and dust-free. If necessary, clean the tip as described on
page 23
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector, dry it with air (e.g. hair blower).