UFM

3030

21

3.07.09

ERR LIMIT

Error limit in % of measured value for plausibility filter on the sensor
paths.

3.07.10

CNT DECR

Counter decrement setting for plausibility filter

3.07.11

CNT LIMIT

Counter limit for plausibility filter
When "0" is set, the plausibility filter will become inactive
Range: 0 through 1000
Factory setting: 0

3.09.00 COMMUNIC Submenu

3.09.00 Communication

3.09.01

PROTOCOL

Communication protocol
OFF (no communication)
HART (HART)
PROFIB PA (PROFIBUS PA)

3.09.02

HART ADDR

HART address
Range: 00 through 16

3.09.03

PP/FF ADDR

PROFIBUS PA/FOUNDATION Fieldbus address
Range: 000 through 126

4.00.00

PARAM ERR

Main menu 4.00.00 Parameter error

4.01.00

FLOW VELOC Volume flow velocity (v) value incorrect. The flow speed is

calculated from the full scale volume flow and the meter size.
Ensure condition 0.5 m/s

≤ v ≤ 20 m/s (1,5 to 66 feet/s) is met!

4.01.01

FULL SCALE

Full-scale value for 100% volume flow rate, see Function 3.01.01

4.01.02

METER SIZE

Meter size, see Function 3.01.08

4.02.00

CURR OUTP

Current output range incorrect. Setting for 100% is compared with
setting for 0%. Ensure condition 100 pct-0 pct

≥ 4 mA is met!

4.02.01

RANGE

Range of current output, see Function 3.04.03

4.02.02

0 pct

Current value for 0% scale, see Function 3.04.04

4.02.03

100 pct

Current value for 100% scale, see Function 3.04.05

4.03.00

LF CUTOFF

Low-flow cut-off range incorrect: If low flow cut-off is set to on, the
value for CUTOFF-OFF is compared with the value of CUTOFF-ON
on.
Ensure condition CUTOFF-OFF – CUTOFF-ON

≥ 1% is met!

4.03.01

LF CUTOFF

Low-flow cut-off, see Function 3.01.05

4.03.02

CUTOFF ON

Cutoff “on” value, see Function 3.01.06

4.03.03

CUTOFF OFF

Cutoff “off” value, see Function 3.01.07

4.04.00

ENERGY

Full scale value for heat energy rate (E) incorrect. The fullscale
value is compared with the maximum value that can be measured
and should meet the condition: Emax < E fullscale < Emax/1000
The maximum value that can be measured is at maximum flow and
200° C temperature difference.

4.04.01

HEAT FS

Full-scale value for 100 % heat energy rate, see Function 3.02.12

4.05.00

PULSE/VOS

Unit of pulse output for velocity of sound function incorrect
Ensure "PULSE RATE" is selected for "VOS"!

4.05.01

PULS FUNCT

Function of pulse output, see Function 3.05.01

4.05.02

PULSE OUTP

Unit of pulse output, see Function 3.05.07

4.06.00

VOS

Velocity of sound range incorrect:
Ensure condition MAX VOS

− MIN VOS ≥ 1 m/s (3.3 feet/sec) is

met!

4.06.01

MIN VOS

Minimum velocity of sound, see Function 3.01.11

4.06.02

MAX VOS

Maximum velocity of sound, see Function 3.01.12

4.07.00

PULSE OUTP

Pulse output frequency value (f) incorrect. The max frequency is
calculated from the pulse/unit setting and the max value of the
measured value.
Ensure condition 1 pulse/hr

≤ f ≤ 2000 pulse/s is met.