3.6

Connection diagram examples

Current output
Active

Ri <= 680 ohm

Passive







For supply:
U = 15…24Vdc, I >= 22mA


Pulse output
Active Passive


R1>= 470 ohm,
R2 = U*R1/(V+ - U)


For supply: U<= 32Vdc, <= 24Vac


Digital input

Analog input

Active

Passive

For supply: U = 15 –
30Vdc, I >= 1,5 mA

4 Start-up

• Check that the flowmeter has been correctly installed.

• With separate systems, check before initial start-up that the correct converter (UFC 030 F) is

used with the correct flow sensor (UFS 3000).

• Order No., see instrument nameplates

• Meter size (DN), Function 3.1.5

• Primary constant GK, Function 3.1.6

• Flow direction, Function 3.1.7

• When powered, the signal converter operates in the measuring mode. TEST, NO ERROR and

IDENT NO. _ _ _ _ _ _ _ of the signal converter appear in succession on the display. This is
followed by display of the actual flow rate and/or the internal count on a continuous or alternating
basis (depending on setting, see Function 3.03 Display or Function 1.02 Display).


UFM

3030

11