KROHNE UFM 3030 EN User Manual
Page 15
UFM
3030
15
1.00.00 OPERATION
Main
menu 1.00.00 Operation
1.01.00
FLOW
Submenu 1.01.00 Flow
1.01.01
FULL SCALE
Full-scale value for 100% volume flow rate, see Function 3.01.01
1.01.02
ZERO VALUE
Zero value, see Function 3.01.02
1.01.03
ZERO CAL
Zero calibration, see Function 3.01.03
1.01.04
MASTER TC
Master time constant, see Function 3.01.04
1.01.05
LF CUTOFF
Low-flow cut-off, see Function 3.01.05
1.01.06
CUTOFF ON
Cut-off active, see Function 3.01.06
1.01.07
CUTOFF OFF
Cut-off de-active, see Function 3.01.07
1.02.00 DISPLAY Submenu 1.02.00 Display
1.02.01 DISP
FLOW Display
of flow, see Function 3.03.01
1.02.02 DISP
TOTAL
Function
of
Totalizer, see Function 3.03.02
1.02.03
TOTAL VOL
Display of Totalizer, see Function 3.03.04
1.03.00 PULSE
OUTP
Submenu 1.03.00 Pulse output
1.03.01
PULSE RATE
Pulse frequency value for 100 % scale, See Function 3.05.08
1.03.02
PULSE/UNIT
Pulse value per volume flow unit, see Function 3.05.09
1.03.03
PULSE/UNIT
Pulse value per energy unit, see function 3.05.10
2.00.00
TEST
Main menu 2.00.00 Test functions
2.01.00 DISPLAY Submenu 2.01.00 Display
2.01.01 DISPLAY Test display, lights all pixels. End with
↵ key
2.02.00 OUTPUTS Submenu 2.02.00 Outputs
2.02.01
CURRENT
Test current output
0 mA
4 mA
12 mA
20 mA
22 mA
Use up arrow to advance. Displayed value directly present at
current output. Actual value present at output after pressing
↵ key.
2.02.02
PULSE
Test pulse/frequency output
1 Hz
10 Hz
100 Hz
1000 Hz
2000 Hz
Use the up arrow to advance. Displayed value directly present at
pulse output. Actual value present at output after pressing
↵ key
2.03.00 INPUTS
Submenu 2.03.00 Inputs
2.03.01
AN INP 1
Test analog input 1
Measure current at analog input 1.
End with
↵ key.
2.03.02
AN INP 2
Test analog input 2
Measure current at analog input 2.
End with
↵ key.
2.03.03
DIG INPUT
Test digital input
Measure level at digital input.
End with
↵ key.
2.03.04
SENSOR
Sensor status, per sensor (6 statuses in total): good, open, short