4 printer check, 5 digital output check, 6 analog output check – KANOMAX 2211 Handheld IAQ Monitor User Manual
Page 54

49
11.4 Printer Check
Problem Possible
Cause(s) / Solution(s)
Refer To
(Page No.)
Printer is not connected properly.
→
Check the connection. Re-connect if necessary.
21
The Baud Rate is not set properly.
→
Check both, CLIMOMASTER and printer, settings.
21
Printer may not be compatible (DPU-H245 and DPU-201GS
are recommended).
Check your printer type.
21
Unable to printout.
Printer connection procedure may not be followed properly.
After a connection is established, you need to turn the IAQ
Monitor power ON, then a printer.
21
Unable to printout the display.
Display is not frozen.
→①
Press
to hold the display.
②
Press
to print out.
21
Unable to cancel the print out.
You cannot cancel the print out.
21
11.5 Digital Output Check
Problem Possible
Cause(s) / Solution(s)
Refer To
(Page No.)
Cable may not be connected properly.
Requires RS-232C cable.
27
Baud rate may not be set properly.
Check baud rate setting on IAQ Monitor and printer.
27
Unable to output data
Communication command may not be correct.
27
11.6 Analog Output Check
Problem Possible
Cause(s) / Solution(s)
Refer To
(Page No.)
Polarity of output terminal may be incorrect.
30
Unable to output data.
Measured value may be in HOLD status.
30
Output appears in tiered pattern.
Output is set per second.
30
Analog output setting may be incorrect.
30
Output value range setting may be incorrect.
30
Output data is incorrect.
Load impedance may be lower than standard value (more than
5kΩ).
30