2 calibration procedure – zero calibration – KANOMAX 2211 Handheld IAQ Monitor User Manual
Page 44

39
8.2 Calibration Procedure – ZERO Calibration
Display
Procedure
Connect ZERO gas tank and IAQ Monitor.
Press
Use
to select [7.CALIBRATION]
Press
.
Use
to select sensor (1.CO2 or 2.CO).
Press
. (CO2 is selected in the example.)
Use
to select [ZERO].
Press
.
Use
to select [1.ZERO], press
.
……
Currently calibrating sensor (CO or CO2)
……
The value before calibration