Storage Recommendations

Prior to using the pH sensor after long-term storage, rinse the sensor with DI water and then soak it in pH 4 buffer
for 1 or 2 hours. This will saturate the glass bulb with hydrogen ions and prepare it for use.

Do not store the pH sensor in DI water because it will deplete the reference solution and drastically reduce the life
of the sensor.

Replacement Parts

Reference Filling Solution (60 mL)................................................................................................................................................0056900

pH Calibration Kit (1 L each pH 4, pH 7, pH 10, DI water)......................................................................................................0032080

Quick Cal (4 x 250 mL).......................................................................................................................................................................0033250

Reference Junction Kit (3 junctions, 60 mL reference filling solution...............................................................................0059620

Storage Solution (500 mL)................................................................................................................................................................0065370

Specifications

Type:

Glass sensing bulb with Ag/AgCl reference half-cell, replaceable single junction,

replaceable

reference

filling

solution

Range:

0 to 12 pH units

Pressure Rating:

211 m (692 ft)

Accuracy:

± 0.1 pH unit

Resolution:

0.01 pH unit

Operating Temperature:

0 to 50° C (32 to 122° F)