Instruction sheet – In-Situ TROLL Net Hub Instruction Sheet User Manual
Page 3

Instruction Sheet
Number
Wire Color
Function
3
Blue
RS485 (+)
4
Red
Power DC+
The center screw terminal block is the master, which
connects to the TROLL
®
Link Telemetry System or
PLC. On 8-port TROLL Net Hubs, which has two circuit
boards, one circuit board master screw terminal is wired
to the second circuit board master screw terminal.
Number
Description
1
Board 1 Master Terminal (to board 2
Master)
2
Board 2 Master Terminal (to telemetry,
PLC, or other TROLL Net Hub)
For bulkhead connector models, disconnect the Red
Power DC+ wire from the screw terminal of any port
that is not being used.
NEVER disconnect the Red Power DC+ wire
from a master terminal.
Internally (9V Battery) Powered 4-port or
8-port TROLL Net Hub
Each screw terminal of an internally powered TROLL
Net Hub is wired the same way, with the exception of
the master terminal(s).
Number
Wire Color
Function
1
Black
Ground
2
Green
RS485(-)
3
Blue
RS485(+)
The center screw terminal block is the master, which
connects to the TROLL
®
Link Telemetry System or
PLC. On 8-port TROLL Net Hubs, which has two circuit
boards, one circuit board master screw terminal is wired
to the second circuit board master screw terminal.
Number
Wire Color
Function
1
Black
To 9V Battery (-)
2
Black
Ground
3
Green
RS485(-)
4
Blue
RS485(+)
5
Red
To 9V Battery (+)
DO NOT connect the red Power DC(+) wire from an
instrument cable to a screw terminal when the
TROLL Net Hub is internally powered. Connecting this
wire causes the instrument to use the 9V battery as an
800-446-7488
3
www.in-situ.com