calibration. User calibration is recommended only if you must conform to a standard
operating procedure or if the conductivity cell has undergone physical change (e.g.,
deposits on conductivity cell walls that cannot be removed or physical damage to the
conductivity cell walls).
1. Fill the calibration cup to the fill line with approximately 130 mL of the desired
calibration solution.
2. Place the calibration cap on the instrument slightly above the restrictor, and
carefully place the instrument in the solution taking care to not force the solution out
the top of the calibration cup.
3. In the Comm Kit Software, select the Sensor Setup tab.
4. Select the Conductivity parameter.
5. Select either 20° C or 25° C as the reference temperature, as indicated by the
reference solution.
6. Click Calibrate.