In-Situ Bulkhead Connector for Use with RuggedCable Systems User Manual
Page 2
(toll-free, US and Canada) or 970 498 1500
www.in-situ.com
Due to continuing product development this information is subject to change without notice. In-Situ and the In-Situ Inc. logo, Win-Situ, TROLL, BaroTROLL, RuggedReader,
RuggedCable, and RDO are trademarks or registered trademarks of In-Situ Inc., Fort Collins, CO, USA. Copyright © 2006, 2008 by In-Situ Inc. All rights reserved.
1 800 446 7488
bULKHEAD cONNEcTOR
Catalog No. 0053240
0053242 rev. 003 03/08
LEvEL TROLL ANALOg (4-20 mA) 2 WiRE
4-20 mA
brown
PLC or Data Logger
GND/ReTuRN
black
+
-
signal
12-36 VDC
LEvEL TROLL / AqUA TROLL sDi-12 3 WiRE
exT PwR
red
GND/ReTuRN
black
Data Logger
9.6-16 VDC
SDI-12
white
RS485 (–)
green
RS485 (+)
blue
AqUA TROLL ANALOg (4-20 mA) 3 WiRE
4-20 mA
brown
PLC or Data Logger
GND/ReTuRN
black
+
-
signal
12-36 VDC
exT PwR
red
MODbUs MAsTER
*
Optional but highly recommended
exT PwR
red
GND/ReTuRN
black
Digital PLC
12-36 VDC
*
RS485 (–)
green
RS485 (+)
blue
with RS485 built in
RS485
RS232
*
Required if port power is not available
DB9F
TD(A)
TD(B)
GND
GND
+12V
RS485 (–)
green
RS485 (+)
blue
pin
5
gnd
pin
3
txd
pin
2
rxd
GND/ReTuRN
black
*
exT PwR
red
*
cONvERTER
Port-Powered RS485 converter,
such as B&B electronics Model
485SD9TB
MODbUs MAsTER
*
Voltage limited by converter
exT PwR
red
GND/ReTuRN
black
Digital PLC
12 VDC
*
RS485 (–)
green
RS232 (RxD)
RS232 (TxD)
Converter
Gnd +12V
RS485 (+)
blue
with RS232 built in (converter required)