DUAL DISPLAY DIGITAL MULTIMETER

USER MANUAL

6

3.SPECIFICATIONS

The specifications are operated under the essential conditions as follows:
z A 1-year calibration cycle.
z An operating temperature of 18 to 28℃ (64.4 to 82.4℉)
z Relative humidity not exceeding 75%

z

Accuracy is expressed as ±(percentage of reading + digits).

z

The AC specification is based on the 50% of duty cycle.

1. DC VOLTAGE

RANGE

RESOLUTION

ACCURACY

INPUT IMPEDANCE

500mV

10μV

10MΩ

5V

100μV

11.1 MΩ

50V

1mV

10.1MΩ

500V

10mV

0.03%+4

10MΩ

1000V

100mV

0.03%+9

10MΩ

Input Impedance

Approx. 10MΩ in parallel with <100pF, all ranges.

Normal Mode Rejection Ratio

>60dB at 60Hz or 50Hz.

Common Mode Rejection Ratio >90dB at 60Hz or 50Hz.
Common Mode Voltage (Max.) 500V dc or peak ac.

Maximum Input

450V dc or peak ac continuous on 500mV range.
1000V dc or peak ac continuous on other range.

dBm (ref 600Ω)

63.80 dBm ~-97.7 dBm

When the input exceeds the full scale of the selected range, the display will appear
“—OL—“ of over-range indication.
2. TRUE RMS AC OR AC+DC VOLTAGE
Accuracy

between 2% of range and full range.

RANGE

20Hz-

45Hz

45Hz-

1kHz

1kHz-

2kHz

2kHz-

10kHz

10kHz-

20kHz

20kHz-

50kHz

500mV

1%+15

0.5%+15

1%+15

2%+30

5%+30

5V

1%+15

0.5%+15

1%+15

2%+30

5%+30

50V

1%+15

0.5%+15

1%+15

2%+30

5%+30

500V

1%+15

0.5%+15

—————————

1000V

1%+15

0.5%+15

—————————

DUAL DISPLAY DIGITAL MULTIMETER

USER MANUAL

7

RANGE

RESOLUTION

INPUT IMPEDANCE

500mV

10μV

10MΩ

5V

100μV

11.1 MΩ

50V

1mV

10.1MΩ

500V

10mV

10MΩ

1000V

100mV

10MΩ

Input Impedance

Approx. 10MΩ in parallel with < 100pF,all ranges.

Maximum Input

450V dc or peak ac continuous on 500mV range.
1000V rms on other range.

dBm ( ref 600Ω)

63.8dBm ~ -97.7dBm

Crest Factor Range 3.0 at full scale.

When the input exceeds the full scale of the selected range, the display will appear
“—OL—“ of over-range indication.
3. FREQUENCY MEASUREMENT AT ACV RANGE

RANGE

FREQUENCY

INPUT LEVEL (SINE WAVE)

ACCURACY

10Hz ~ 50kHz

≧ 120mV

0.05%+1

500mV

50k ~ 150kHz

≧ 200mV

0.05%+1

5V

10Hz ~ 200kHz

≧ 1.2V

0.05%+1

50V

20Hz ~ 200kHz

≧ 1.2V

0.05%+1

500V

20Hz ~ 1kHz

≧ 12V

0.05%+1

AC+DC measurement does not support AC+Hz function.

Maximum Input

450V peak ac continuous on 500mV range.
500V peak ac continuous on the other range.

4. DC Current

RANGE

RESOLUTION

ACCURACY

BURDEN VOLTAGE

500μA

0.01μA

0.2%+2

0.7Vmax.

5mA

0.1μA

0.2%+2

0.7Vmax.

50mA

1μA

0.2%+2

0.7Vmax.

500mA

10μA

0.2%+2

0.8Vmax.

2A

100μA

0.3%+2

0.8Vmax.

20A

1mA

0.3%+2

0.9Vmax.