14

Note

l

If the value of voltage to be measured is unknown, use the maximum measurement

position (500V) and reduce the range step by step until a satisfactory reading
is obtained.

l

The LCD displays “1” indicating the existing selected range is overload; it is

required to select a higher range in order to obtain a correct reading.

l

In each range, the Meter has an input impedance of approx. 10M

. This loading

effect can cause measurement errors in high impedance circuits. If the circuit
impedance is less than or equal to 10k

, the error is negligible (0.1% or less).

l

When DC voltage measurement has been completed, disconnect the connection

between the testing leads and the circuit under test.

Model GDM-350A: OPERATING MANUAL

Measurement Operation (2)