26

DC Current

Range

Resolution

Accuracy

2000µA

1µA

±(1% Reading +2 Digits)

20mA

0.01mA

±(1% Reading + 2 Digits)

200mA

0.1mA

±(1.2% Reading +2 Digits)

10A

0.01A

±(2% Reading +5 Digits)

Overload Protection:
mA range: F2 fuseφ6×25mm, F 1A H 240V (CE)
10A range: F1 fuseφ6×25mm, F 10A H 240V (CE)

Warning:

When ≤ 5A Continuous measurement is allowed.
When > 5A Measurements must not take longer than 10 seconds with a wait of 15
minutes between measurements.