16

A

PENDIX

GDP Series Specifications

GDP-025

Bandwidth

DC-25MHz (-3dB)for x50, x200
DC-50MHz (for attenuation x20)

Attenuation

x20, x50, x200

Accuracy

± 2%

Voltage

Input

Range (DC+AC
peak to peak)

≤ 140Vp-p for x20 (؆ 45Vrms)
≤ 350Vp-p for x50 (؆ 110Vrms)
≤ 1400Vp-p for x100 (؆ 450Vrms)

Permitted Max
Input Voltage

Max differential voltage: 1400 V (DC+AC
peak to peak) or 450 Vrms
Max voltage between each input terminal
and ground: 600 Vrms

Input

Impedance Differential: 4MΩ/1.2pF

Between terminals and ground:
2MΩ/2.3pF

Output

≤ ±6.5V

Output

Impedance

50Ω

Rise Time

14ns for x50, and x200; 23.4ns for x20

Rejection Rate on
Common Mode

60Hz: >80dB; 100Hz: >60dB; 1MHz:
>50dB

Power Supply

External 9V DC power supply

Consumption

200mA

max

Environmental
Operating
Conditions

Indoor Use only.
Temperature: Reference: +20˚C to
+30˚C; Use: 0˚C to +50˚C; Storage: -30˚C
to +70˚C
Relative Humidity: Reference: ≤70%;
Use: 10% to 85%: 10% to 90%
Storage: 10% to 90%

Dimensions

and

Weight

195mm x 55mm x 30mm: 250g