Poor calibration repeatability, Unstable output/ sudden spiking, Nuisance alarms – Detcon CXT-DM User Manual

Page 42: Transmitter not responding

background image

Model CXT-DM

• Replace the plug-in toxic sensor.

Poor Calibration Repeatability

Probable Causes: Failed Sensor, use of wrong Cal Gas or problems w/ cal gas and delivery, Interference Gases

• Check validity of span gas with regulator and sample tubing in place using pull tube or other means

(check MFG date on cal gas cylinder).

• Use proper cal gas regulators and tubing for highly corrosive gases (HF, HCI, Cl2, NH3, HBR, F2,

etc.)

• Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded). H2S sensors assemblies use CXT-DM Series Splashguard Adapter Assembly with integral
filter. Clean or replace if necessary.

• Replace the plug-in toxic sensor.

Unstable Output/ sudden spiking

Possible Causes: Unstable power supply, inadequate grounding, or inadequate RFI protection.

• Verify Power source is stable.

• Verify field wiring is properly shielded and grounded.

• Contact Detcon to optimize shielding and grounding.

• Add Detcon’s RFI Protection Circuit accessory if problem is proven RFI induced.

Nuisance Alarms

• Check condulet for accumulated water and abnormal corrosion on terminal board.

• If nuisance alarms are happening at night, suspect condensation in condulet.

• Add or replace Detcon’s Condensation Prevention Packet P/N 960-202200-000.

• Investigate the presence of other target gases that are causing cross-interference signals.

• Determine if cause is RFI induced.

Transmitter not responding

• Verify condulet has no accumulated water or abnormal corrosion.

• Verify required batteries are installed and have enough charge to power the sensor.

• Swap with a known-good ITM to determine if ITM is faulty.


Low Battery / Low Voltage Fault

Possible Cause: Lithium Ion batteries are depleted.

• Change out all three of the Lithium Ion non-rechargeable ‘C’ size batteries with new batteries.

NOTE

Contact the Detcon Service Department for further troubleshooting assistance at

281-367-4100.