Model CX-DM
Figure 11 Magnetic Programming Switches
NOTE
While in Program Mode, if there is no magnetic switch interaction after 4 consecutive menu
scrolls, the sensor will automatically revert to normal operating condition. While changing
values inside menu items, if there is no magnet activity after 3 to 4 seconds the sensor
will revert to the menu scroll. If the sensor is in Bump Test mode, the display will remain
active.
3.1
Operator Interface
The operating interface is menu-driven via the two magnetic program switches located under the target marks
of the sensor housing. The two switches are referred to as PGM1 and PGM2. The menu list consists of three
menu items that include sub-menus:
Normal Operation
Concentration reading is displayed once every 10 seconds.
Calibration Mode
AutoZero
AutoSpan
Program Mode
View Sensor Status
CX-DM ##.##
Range ###
Autospan Level ##
Modbus ID ##
Tempcomp Level ##
Last Cal ## Days
Sensor Life ###%
Temperature ##C
Set Detection Range
Program Switch #1
LED Display
Program Switch #2
PGM2
SPAN
PGM1
ZERO
MODEL
CX-DM
detcon inc.
H
2
S Sensor
CX-DM Instruction Manual
Rev. 1.3
Page 12 of 40