Detcon FP-524D-HRT User Manual
Page 50
Model FP-524D-HRT
FP-524D-HRT Instruction Manual
Rev. 0.1
Page 46 of 48
1000 ohms maximum loop load @ 24VDC
0mA
All Fault Diagnostics
2mA
In-Calibration
4-20mA
0-100% full-scale
22mA
Over-range condition
HART Version
7.0
Manufacture ID
0x6007
Device ID
0xE08B
Status Indicators:
4-digit LED Display with gas concentration, full-script menu
Prompts for AutoSpan, AutoZero, Set-up Options, and Fault Reporting
Faults Monitored:
Loop Fault, Input Voltage Fault, Zero Fault, Sensor Fault,
Processor Fault, Memory Fault, Calibration Fault(s)
Cable Requirements:
Power/Analog: 3-wire shielded cable
Maximum distance is 13,300 feet with 14 AWG
Mechanical Specifications
Length:
5 inches (200mm), including Splashguard
Width:
6 inches (140mm)
Weight:
2.7lbs (3.4Kg)
Mechanical Connection:
¾” Male NPT threaded connection
Electrical Connection:
three 14 gauge (maximum) wire terminal landings