Model FP-524D-HRT

FP-524D-HRT Instruction Manual

Rev. 0.1

Page 14 of 48

Software Flowchart

dec

PGM1/2 (M)

mA Output XX.XX

Sensor Temp XX C

Voltage XX.X VDC

Gas Factor

Sensor Life XXX%

Bridge Current

Bridge Voltage

PGM1/2 (3)

PGM1 (3)

PGM1 (3)

PGM2 (3)

Version X.XX

AutoSpan @ XX

Range XXX ppm

Last Cal XX Days

inc

Auto Time-Out

View Sensor Status

PGM1/2 (3)

PGM1/2 (M)

Model Type

PGM2 (3)

PGM1 (S)

PGM1/2 (3)

PGM2 (S)

Auto Time-Out

Set Cal Factor

##

AutoTime-out

PGM1/2 (3)

PGM1/2 (M)

Set AutoSpan Level

Calibration Mode

(Auto Span)

Normal Operation

PGM1/2 (M)

PGM1/2 (M)

Simulation

LEGEND:

PGM1 - Program Switch Location #1
PGM2 - Program Switch Location #2

(S) - Momentary Swipe
(M) - Momentary hold of Magnet during text

scroll until the ">" appears, then release

(3) - 3 second hold from ">" prompt
(10) - 10 second hold from ">" prompt

Auto Time-out - 5 seconds

inc - Increase
dec - Decrease

#, ##, ### - numeric values

PGM2 (10)

PGM1/2 (10)

Signal Output Check

Auto Time-Out

Auto Time-Out

Auto Time-Out

Set Bridge Voltage

inc

PGM1/2 (3)

PGM2 (S)

PGM1 (S)

#.##

PGM1/2 (3)

PGM1/2 (M)

Set Gas Factor

dec

PGM1/2 (M)

Defaults Restored

PGM1/2 (10)

Restore Defaults

Auto Time-Out

PGM1/2 (3)

inc

PGM1/2 (3)

PGM2 (S)

PGM1 (S)

##

dec

inc

PGM1/2 (3)

PGM2 (S)

PGM1 (S)

#.##

dec

Calibration Mode

(Auto Zero)

Cal Factor

Figure 14 FP-524D-HRT Software Flowchart

3.3 Normal Operation

In normal operation, the display continuously shows the current sensor reading, which will normally appear as
“0”. Once every minute, the LED display will flash the sensor’s units of measure and the gas type (i.e.”%
LEL”). If the sensor is actively experiencing any diagnostic faults, a “Fault Detected” message will scroll
across the display on the display once every minute instead of the units of measure and the gas type. At any
time, while the sensor is in “Fault Detected” mode, PGM1 or PGM2 can be swiped to prompt the sensor to
display a list of the active faults.

In normal operation, the 4-20mA current output linearly corresponds with the full-scale range.