Model FP-524D-HRT

FP-524D-HRT Instruction Manual

Rev. 0.1

Page 32 of 48

Figure 18 FP Configuration Setup

The Range Setup will display the Model Type of the sensor, followed by the PV LRV (Primary Variable
Lower Range Value) and the PV URV (Primary Variable Upper Range Value), and the Conc Units
(Concentration Units), the display may also show the Sensor Range. The Sensor Setup portion of the screen
will display sensor specific parameters:

 Gas Factor

 Cal Factor

 Bridge Current

 Bridge Voltage

 Sensor Life

NOTE: The values above are read when the HRT Bridge boots up and are not updated
in real-time.

4.4.3.2 Calibration

The Calibration screen displays the days since the last calibration, and the auto span level. This screen also
allows the user to calibrate the sensor by performing an Auto Zero Calibration and an Auto Span Calibration.
Calibration of the sensor using this feature also notifies the Host that the sensor is in calibration mode.