TO RESET THE METER TO DEFAULT SETTINGS IN BASIC MODE:
1. Use the
keys from the BASIC main menu to highlight #4: DEFAULTS.
Press the SELECT key. The meter will display the following question:
2. Use the
keys to answer “YES” or “NO.” If you answer “yes,” the meter
will return to factory default settings in basic mode; i.e. Species = Douglas Fir,
Temperature correction = 70 F, Alarm = OFF in pin mode. In Scan mode, Alarm
= “off.”
USING THE METER IN ENHANCED MODE:
To use TotalCheck in enhanced mode, use the
keys to highlight #5: ENHANCED.
Setting the meter to enhanced mode requires you to create job definitions and store
readings as described below.
NOTE: Although TotalCheck’s default mode is BASIC, it will remember the mode
last saved at power off, and will remain in the saved mode at next power up.
Settings stored in each particular mode will not be affected by restoring defaults
in a different mode.
JOB DEFINITION:
Using TotalCheck in ENHANCED mode allows you to define jobs. In this mode,
TotalCheck also stores each reading and allows you to download the data to a PC or
laptop via Infrared connection. Jobs may be defined only when using the meter in pin
mode or RH-T mode. Scanned readings cannot be part of a job.
TO CREATE OR MODIFY A JOB IN PIN MODE:
1. From the Enhanced Main Menu, highlight PIN MODE.
2. Use the
keys to highlight JOB. Press SELECT to enter the job sub
menu.