Menu operation – DeFelsko PosiTector UTG v.2.2 User Manual
Page 7
7
Menu Operation
Menu Operation
Gage functions are menu controlled. To access the
Menu, turn the Gage on, then press the button.
3. Select the Zero menu option
(LCD will display image shown
in Figure A). Hold the probe in
the air and press the
button
to select OK to continue.
Figure A
4. When the image shown in
Figure B is displayed, press
and hold the probe against the
zero plate.
5. When the Gage beeps and the
image shown in Figure C
appears, remove the probe
from the zero plate.
Figure C
Figure B
The Gage will display 0.00.
DO NOT place probe on the Zero plate
until Step 4.
Important:
1. Make sure the Gage is on and the probe is
wiped clean.
2. Apply a single drop of couplant onto the zero
plate located on the underside of the Gage. DO
NOT apply couplant directly onto the probe face.
Current selection is displayed
with darkened background
Reset
Gage Info
Flip Display
Units
Set Clock
Language
Memory
Zero
Cal Settings
SE Mode
✓
HiLo Alarm
Setup
Some menu items have a tick box
to their right to indicate current
status. An empty box indicates
that feature in not active.