2. Technical Data
2.1
Humidity Detection
Ranges are possible from 0-100 %RH
as well as absolute Humidities, Dewpoints
and Enthalpie, programed is a range of
...........................
Resolution
0,1 % RH
Sensing Element (capacity type)
SE – 02
Influence of Temperature on the Humidity
± 0,01 % RH / °C
Repeatability
< 0,5 % RH
Hysteresis for 4 hours Cycle
10 % - 95 % - 10 % RH
< 0,5 % RH
Unlinearity of Electronics / Humidity
< 1 % RH
2.2
Temperature Detection
Ranges are possible from –50°C to
+150°C, programed range is
..................°C
Resolution
0,1 °C
Temperature sensing Element
Pt - 1000
System Accuracy of Temperature
± 0,35 °C
Unlinearity of Electronics / Temperatur
< 0,15 °C
2.3
Power Supply
Standard Power Supply
90-240 VAC,
47-63 Hz
or Low Voltage Supply
10-30 VDC/9-25 VAC
2.4
Maximum Operational Temperatures
Transmitter Electronics
-20....+40 °C
Probe Electronics
-40....+60 °C
Probe Tube with Sensors
-40....+120 °C
2.5
Power Consumption
Max. Power Consumption
approx. 5 VA
HYGROCONTROL Manual for Type 86 Ex Page
5