The display now shows the different submenus you may choose:

5.1 Calibration Data

Calib-Data

With the buttons under the arrows you may choose the function you want and with
-button the submenu is opened. You will find the different submenus as follows:

HYGROCONTROL Manual for Type 86 Ex Page

5. Operating the Instrument with Display-Option

From normal display of Humidity and Temperature you get into the menus for
Configuration and Calibration by pressing one of the 4 buttons on the display-
board.

26

-- SETTINGS --
List Calibration Data
List Configurat. Data
Calibrate (PW)
Cofigure (PW)
Display Setup

4

4

F freq: 51732
R freq: 51217

F freq: 51732
R freq: 51217

= brings the next value into the display

= also brings the next value into the display

= Stops display of calibration data

= encreasing as long as sensors frequency is detected

= actual Frequency of the Humidity sensor
= actual Frequency of the Reference

Setup: 0.0%RH
F diff: 1395
Rec.#: 0 ReCal:101

Any pressing of the “arrow”-buttons brings the next
calibration point and the corresponding difference
frequency into the display

Rec.# = Number of values in the memory,
20 and 21 are first 2 values of factory calibration
ReCal: 101 means: 1. Digit = deleted by customer
2. and 3. Digit = 1x recalibrated by customer

This menu shows stored values of calibration data, but values cannot be changed!

= letters are white on back background
and this is the submenu choosen.

(PW) = for these submenus a Password is neccessary.

Calib-Data