Table 2

„P”„K” „P”

LED „H3” is blinking Quickly

Delete all Calibration Values

F

no action

D-E

„K”

Adjust to Value according to

your Reference Temperature

Calibration of Temperature

high“ (not yet realized)

C

„K”

Adjust to Value according to

your Reference Temperature

Calibration of Temperature

low

B

„K”

See Table 3

Adjustment of Dimension

Temperature

A

„K”

See Table 3

Adjustment of Dimension

Humidity

9

„K” 2)

Temperature-max. Value

Humidity stays as before

Adjustment of Temperature

Range at upper Limit

8

„K” 2)

Temperature-min. Value
Humidity stays as before

Adjustment of Temperature

Range at lower Limit

7

„K” 2)

Humidity-max. Value

Temperature stays as before

Adjustment of Humidity

Range at upper Limit

6

„K” 2)

Humidity-min. Value

Temperature stays as before

Adjustment of Humidity

Range at lower Limit

5

„K” 1)

Temperature-max. Output

Humidity stays as before

Adjustment of Temperature

Output at upper Limit

4

„K” 1)

Temperature-min. Output

Humidity stays as before

Adjustment of Temperature

Output at lower Limit

3

„K” 1)

Humidity-max. Output

Temperature stays as before

Adjustment of Humidity

Output at upper Limit

2

„K” 1)

Humidity-min. Output

Temperature stays as before

Adjustment of Humidity

Output at lower Limit

1

"M"

Measuring Values

Measuring Mode

0

Jumper

K M P

Signal Output

Activated Function

Code Switch

„S1” Position

Note 1) The corresponding output is encreased as long as the code switch

„S1” is turned slowly to right (LED „H2“ is lightened), and is lowered
as long as the code switch „S1” is turned slowly to left (LED „H4” is
lightened). Values are stored if code bridge „X4” is pulled out of
Position "K". During storage all 3 LED’s are lightened.

Note 2)

If you adjust the lower limits of the ranges of Humidity or
Temperature, 0% of the corresponding output is equivalent a value of
-100, 50% of the corresponding output is the equivalent for a value of
0, and 100% of the corresponding output is the equivalent for a value
of 100. (Example: the corresponding output may be set to 4 to 20 mA
- and the ranges should be set to 0 to 100%rF und –50 to +150°C,
then the lower limit 0%RH is set by an output of 12 mA and the lower
limit –50°C is set by an output of 8 mA).

If you adjust the upper limits of the ranges for Humidity or
Temperature, 0% of the corresponding output is equivalent a value of
0, 50% of the corresponding output is the equivalent for a value of
100, and 100% of the corresponding output is the equivalent for a
value of 200. (Example: the corresponding output may be set to 4 to

HYGROCONTROL Manual for Type 86 Ex Page

20