Hygrocontrol manual for type 86 ex page – COSA Xentaur Hygrocontrol Type 86 Ex User Manual
Page 21
Table 2
„P”„K” „P”
LED „H3” is blinking Quickly
Delete all Calibration Values
F
no action
D-E
„K”
Adjust to Value according to
your Reference Temperature
Calibration of Temperature
„high“ (not yet realized)
C
„K”
Adjust to Value according to
your Reference Temperature
Calibration of Temperature
„low“
B
„K”
See Table 3
Adjustment of Dimension
Temperature
A
„K”
See Table 3
Adjustment of Dimension
Humidity
9
„K” 2)
Temperature-max. Value
Humidity stays as before
Adjustment of Temperature
Range at upper Limit
8
„K” 2)
Temperature-min. Value
Humidity stays as before
Adjustment of Temperature
Range at lower Limit
7
„K” 2)
Humidity-max. Value
Temperature stays as before
Adjustment of Humidity
Range at upper Limit
6
„K” 2)
Humidity-min. Value
Temperature stays as before
Adjustment of Humidity
Range at lower Limit
5
„K” 1)
Temperature-max. Output
Humidity stays as before
Adjustment of Temperature
Output at upper Limit
4
„K” 1)
Temperature-min. Output
Humidity stays as before
Adjustment of Temperature
Output at lower Limit
3
„K” 1)
Humidity-max. Output
Temperature stays as before
Adjustment of Humidity
Output at upper Limit
2
„K” 1)
Humidity-min. Output
Temperature stays as before
Adjustment of Humidity
Output at lower Limit
1
"M"
Measuring Values
Measuring Mode
0
Jumper
K M P
Signal Output
Activated Function
Code Switch
„S1” Position
Note 1) The corresponding output is encreased as long as the code switch
„S1” is turned slowly to right (LED „H2“ is lightened), and is lowered
as long as the code switch „S1” is turned slowly to left (LED „H4” is
lightened). Values are stored if code bridge „X4” is pulled out of
Position "K". During storage all 3 LED’s are lightened.
Note 2)
If you adjust the lower limits of the ranges of Humidity or
Temperature, 0% of the corresponding output is equivalent a value of
-100, 50% of the corresponding output is the equivalent for a value of
0, and 100% of the corresponding output is the equivalent for a value
of 100. (Example: the corresponding output may be set to 4 to 20 mA
- and the ranges should be set to 0 to 100%rF und –50 to +150°C,
then the lower limit 0%RH is set by an output of 12 mA and the lower
limit –50°C is set by an output of 8 mA).
If you adjust the upper limits of the ranges for Humidity or
Temperature, 0% of the corresponding output is equivalent a value of
0, 50% of the corresponding output is the equivalent for a value of
100, and 100% of the corresponding output is the equivalent for a
value of 200. (Example: the corresponding output may be set to 4 to
HYGROCONTROL Manual for Type 86 Ex Page
20