If there was choosen Calibrate (PW) in the main menu
and was acknowledged by button , the Password
must now be typed in. You get the first 2 digits of the
4-digit Password by the sum of the first 3 digits of the
Serial number of the Transmitter. (Leading 0 must be
encluded). Last 2 digits of the Password you will get
by the sum of the last 2 digits of the Serial number of
the Transmitter. (Leading 0 must be encluded)

5.3 Calibration

After the suggested fixed Humidities from 0% to 95%RF
you may also choose “Other Humidity”, “Delete Memory”
and “Ambient Temp”.
If you choose “Other Humidity” you may change the value
of the blinking digit in the same way as you did with the
Password. If the value of the Humidity is the one you
want, you get into the calibration by button .

-- Calibrate --

- Kalibrieren -

Calibrate: 50

act H: 47.3 rH

f diff: 2615

act T: 24,5°C

Password
0000

Setpoint 0%RF
Setpoint 10%RF
Setpoint 20%RF
Setpoint 35%RF

At least you will see the Ser.-No. of the Transmitter

The blinking digit may be encreased by button . Button switches over to the
next digit. If the Password is completed, button brings you into the calibration menu.

Example: Ser.-No. Transmitter = 30453 results in Password 0708

Config-Data

Serial number:
= 00000

Ok!

= this is the Humidity value to be calbrated

= Humidity value shown with the old calibration memory.

= actual value of Temperature

button ok keeps the actual value of F diff. steady and
button will write this as a new value into the memory.

The display will now show for example:

If your choice was “ambient Temp” the display shows after the acknowledgement:

HYGROCONTROL Manual for Type 86 Ex Page