3.2.5 Calibration

Procedure

Proceed as follows:

- Mount the calibration chamber as described before
- Remove the bottom of the chamber to fill it with the desired solution
- Break the neck of the ... %RH ampoule and fill its content into the bottom
- Close the calibration chamber thoroughly in vertical position to avoid

contact of the liquid with the sensor

- Place code switch "S1" to a position according to the choosen value given

in table 1 and code bridge "X4" to position "P"

Table 1

brown

magenta

green

yellow

red

blue

white

Granulate

Colour Code of

Calibration Solution

95

80

65

50

35

20

10

0

Calibration Solution

(% RH)

7

6

5

4

3

2

1

0

Position of Code

Switch „S1”

- The LED „H3" starts blinking now slowly as long as unstable humidities are

detected - the display (if present) shows the actual value of humidity and
the humidity to be calibrated

- After stability of humidity is detected, the LED „H3” and the display (if

present) show a steady signal - the ... %RH calbration is finished and
may be stored.

- If you now remove the code bridge „X4” out of position „P”, the new value

will be stored and this is shown by the lightning of LED’s „H2”, „H3” and
„H4”.

- If you don’t want to store the new value, you must turn the code switch „S1”

to another position before removing the code bridge „X4”

- Place code switch „S1” into position "0" - LED will turn dark and the

humidity output will show the calibrated ... %RH value with a deviation of
max. 0,5%RH.

- Place code bridge „X4" into position „M" now, to return to normal measuring

mode.

If more calibrations are to follow, proceed as described before, using the
solution you want to calibrate with.

Never forget to clean the bottom of the chamber and dry it
thoroughly before starting the next calibrating procedure!