Hygrocontrol manual for type 86 ex page – COSA Xentaur Hygrocontrol Type 86 Ex User Manual
Page 25
Table 3: Display at 100% RH over Ice at different Temperatures
75,5
0,37
0,49
- 30
79,0
0,64
0,81
- 25
81,7
1,03
1,26
- 20
86,9
1,66
1,91
- 15
90,6
2,60
2,87
- 10
95,3
4,02
4,22
- 5
100
6,11
6,11
0
M (%RH)
PSE (mbar)
PSF (mbar)
Temperatur (°C)
PSF = Saturated Vapor Pressure over Chilled Water
PSE = Saturated Vapor Pressure over Ice
M
= Rel. Humidity displayed by HYGROMESS Instruments.
4.7 Measurements under Pressure or Vaccuum
All probes can be used at pressures of 0,9 to 1,3 bar. Only special probes
with a glas feed-through (which predicts SS - 316 tube material) may be
used under pressures of 0,03 to 30 bar. As rel. Humidity is directly
proportional to the pressure (f.E.: increase of pressure by factor 2 also
increases rel. Humidity by factor 2, when temperature and content of water
remain constant!) you shoud take care of a place with constant pressure
values, when mounting the probe. Be aware that pressure may change,
when diameters of tubes change - and air streams may be turbulent at
these places.
HYGROCONTROL Manual for Type 86 Ex Page
24