Hygrocontrol manual for type 86 ex page – COSA Xentaur Hygrocontrol Type 86 Ex User Manual
Page 13
The current outputs may be set to be 0-20 mA or 4-20 mA. Normally your
desired version is programed by the factory - if you want to change it
please see chapter „CONFIGURATION” where you find how to do this.
3.1.2
Tables of analogue outputs
Following tables show some examples of relations between outputs and
ranges for temperature and relative humidity.
3.1.2.1 Examples for outputs of relativ humidities (0 - 100 %)
10,00
5,00
2,00
1,00
20,00
20,00
100
9,50
4,75
1,90
0,95
19,20
19,00
95
9,00
4,50
1,80
0,90
18,40
18,00
90
8,50
4,25
1,70
0,85
17,60
17,00
85
8,00
4,00
1,60
0,80
16,80
16,00
80
7,50
3,75
1,50
0,75
16,00
15,00
75
7,00
3,50
1,40
0,70
15,20
14,00
70
6,50
3,25
1,30
0,65
14,40
13,00
65
6,00
3,00
1,20
0,60
13,60
12,00
60
5,50
2,75
1,10
0,55
12,80
11,00
55
5,00
2,50
1,00
0,50
12,00
10,00
50
4,50
2,25
0,90
0,45
11,20
9,00
45
4,00
2,00
0,80
0,40
10,40
8,00
40
3,50
1,75
0,70
0,35
9,60
7,00
35
3,00
1,50
0,60
0,30
8,80
6,00
30
2,50
1,25
0,50
0,25
8,00
5,00
25
2,00
1,00
0,40
0,20
7,20
4,00
20
1,50
0,75
0,30
0,15
6,40
3,00
15
1,00
0,50
0,20
0,10
5,60
2,00
10
0,50
0,25
0,10
0,05
4,80
1,00
5
0,00
0,00
0,00
0,00
4,00
0,00
0
0 ... +10
V
0 ...+5
V
0 ... +2
V
0 ... +1
V
4 ... 20
mA
0 ... 20
mA
% r. F.