LumaSense Technologies IS 6 Advanced User Manual
Page 36

IS 6 Advanced / IGA 6 Advanced Manual
Data format UPP
Ò
· 36
Description
Command
Parameters
Response Time t
90
AAezX
X=0 to 9
0=min
1=1 ms 2=3 ms
3=5 ms
4=10 ms
5=50 ms
6=250 ms
7=1 s
8=3 s
9=10 s
Temp Display °C or °F AAfhX
Output: X=0 Display
in
°C
X=1
Display in °F
Device address
AAgaXX
XX=(00 to 97)
00 to 97=regular device addresses
99=global address with response
Internal temperature
(read)
AAgt
AAtm
Answer: DDD 3 decimal digits (000 to 098 °C or
032 to 210 °F)
gt=current temp.
tm=maximum temp. (memory)
Laser
AAlaX
AAla
X=0 off X=1 on
Answer: 1-digit: “0” or “1”
Software simulation
of external clearance
AAlx
Clears maximum storage
Clear Peak Memory
t
clear
AAlzX
X=0 bis 9
0=AUS
1=10
ms
2=50
ms
3=250
ms
4=1 s 5=5 s
6=25 s
7=EXTERN
8=AUTO
9=HOLD
Basic
range
(read)
AAmb
Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Sub
range
(read)
AAme
Answer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Sub range (set)
1. AAm1XXXXYYYY
XXXXYYYY=2x4 hex-digit for lower and upper sub range
limit (°C or °F)
2. AAm2
AAm2 confirms the change (auto reset)
Measuring
value
AAms
Answer: QQQQQ (88880=Overflow)
5 decimal digit (in °C or °F, last digit is 1/10 °C or °F)
Device type
AAna
Output: “IS 6 Advanced / IGA 6 Advanced”
(16 ASCII-characters)
Read
Parameters
AApa
Answer: 15 decimal digits
DD.............
: Emissivity (see em)
..D............
: Response time (see ez)
...D...........
: Clear peak memory (see lz)
....D..........
: Analog output (see as)
.....DD........
: Internal temperature (see gt)
.......DD......
: Device address (see ga)
.........D.....
: Baud rate (see br )
..........01000
: Fixed value
Serial number
AAsn
Output: XXXXX (hex 5-digit)