5.4 Basic settings
All preset values for the
device can be displayed
and modified, if necessary under
the Devices/Parameters
window.
Changing an existing pyrometer
setting can be accomplished by
typing a value in an input box or
by selecting a preset value from
the list field. Choose the correct
settings for your application
from the displayed options.
This window contains the
parameter settings described in
Chapter
, Parameters.
5.4.1 Open/Save
The open and save buttons enable you to save customized configurations, which can
then be opened and reused at any time.
5.4.2 1 measure…
The 1 measur… function shows the current measuring temperature in the
pyrometer parameters window for approximately 1 second.
5.4.3 Print
The print icon will generate a screenshot of the parameter form and allow you to
send it a printer.
5.4.4 Close
The close icon will close the pyrometer parameters window.
5.4.5 Test
Clicking on the Test
icon will open a
window that allows you to
directly communicate with the
pyrometer using the UPP
(Universal Pyrometer Protocol)
Data Format commands.
After entering an interface
command (00 is the adjusted
address ex works, ms is the
command reading
temperature value) and
clicking on Send, the window
shown to the right will be
opened.
This window already shows the answer of the pyrometer in 1/10°. The actual temperature
reading is 599.0 °C.